10.5.2 Installing Gas Cylinders

Each ZO21S Standard Gas Unit comes with six zero gas cylinders including a spare. Each

gas cylinder contains 7-liters of gas with a 0.95 to 1.0 vol%O

2

(concentration varies with each

cylinder) and nitrogen, at a pressure of 700 kPaG (at 35 °C).
The operating details and handling precautions are also printed on the product. Please read them

beforehand.
To install the gas cylinder, follow these steps:
(1) Attach the zero gas valves onto the gas cylinder. First, turn the valve regulator of the zero

gas valves counterclockwise to completely retract the needle at the top from the gasket

surface. Maintaining the valve in this position, screw the valve mounting into the mouthpiece

of the gas cylinder. (If screw connection is proper, you can turn the screw manually. Do not

use any tool.) When the gasket comes in contact with the mouthpiece of the gas cylinder

and you can no longer turn it manually, tighten the lock nut with a wrench.

(2) Remove the carrying case from the standard gas unit. The case is attached to the unit with

six screws. So, loosen the screws and lift them off.

(3) Slide the gas cylinder through the hole in the back of the unit and connect the tube (the

piping in the unit) to the valve connections. Insert each tube at least 10 mm to prevent

leakage, and secure it using a tube clamp.

(4) Attach the gas cylinder to the case. Extend the valve regulator of the zero gas valves

through the hole in the front panel of the unit and secure the bottom of the cylinder with the

clamp.

(5) Take note of the oxygen concentration of the sealed gas indicated on the gas cylinder and

replace the carrying case. Enter the oxygen concentration of the sealed gas, following

instructions on the converter display. Also check that no piping is disconnected.

Thus, the work of installing a gas cylinder is completed. However, gases in the cylinders cannot

immediately flow out after these procedures. To discharge the gases, it is necessary for the

needle in the zero gas valves to puncture a hole in the gas cylinder (see Section 10.5.3).

10.5.3 Calibration Gas Flow

<Preparation before calibration>
(1) To operate the standard gas unit, place it on a nearly horizontal surface in order to allow the

flow check to indicate the precise flow rate. In addition, a power supply for driving the span

gas (air) supply pump is required near the unit (the length of the power cord attached to the

unit is 2 m). Select a suitable location for the unit near the installation site of the converter.

(2) Connect the tube connector port of the standard gas unit to the calibration gas inlet of the

detector, using a polyethylene resin tube with an outside diameter of 6 mm. Be careful to

prevent gas leakage.

(3) Fully open the stop valve mounted on the calibration gas inlet of the detector.
(4) Enter the oxygen concentration of the sealed gas (noted from the cylinder) into the

converter. Also check that the oxygen concentration of the span gas is correctly set

(21 vol%O

2

for clean air). When using the ZO21S Standard Gas Unit (for use of the

atmospheric air as a span gas), use a hand-held oxygen analyzer to measure the actual

oxygen concentration, and then enter it.