<8. Detailed Data Setting>

8-11

IM 11M12A01-02E

Table 8.5 Output Contact Settings

Item to be selected

Brief description

Alarm

and

Error

settings

High-high limit alarm

If “High-High alarm ON” is selected, contact output occurs when the

high-high limit is issued. To do this, it is required, in Alarms setup, that

the high-high alarm be set on beforehand (see Section 8.3).

High limit alarm

If “High alarm ON” is selected, contact output occurs when the high

limit alarm is provided. To do this, it is required, in Alarms setup, that

the high limit alarm be set on beforehand (see Section 8.3).

Low limit alarm

If “Low alarm ON” is selected, contact output occurs when the low

limit alarm is provided. To do this, it is required, in Alarms setup, that

the low limit alarm be set on beforehand (see Section 8.3).

Low-low limit alarm

If “Low-Low alarm ON” is selected, contact output occurs when the

low-low limit alarm is issued. To do this, it is required, in Alarms setup,

that the low-low alarm be set on beforehand (see Section 8.3).

Calibration coefficient alarm If Calibration coefficient alarm is ON (enabled), then when a zero

calibration coefficient alarm (Alarm 6) or span calibration coefficient

alarm (Alarm 7) occurs then calibration coefficient alarm contact

output occurs (see Sec. 12.2.1)

Startup power stabilization

timeout alarm

If set ON then contact output occurs when startup power stabilization

timeout alarm (Alarm 8) occurs (see Sec.12.2.1)

Error

If “Error ON” is selected, contact output occurs when an error is

issued. (See Chapter 12, Troubleshooting).

Other

settings

Warm-up

If “Warm-up ON” is selected, contact output occurs during warm-up.

For the definition of Warm-up (see Section 8.2.1).

Output range change

If “Range change ON” is selected, contact output occurs (“answer-

back signal to a range change signal”) while a range change signal is

applied to a contact input. To do this, it is required, in Input contacts

setup, that the range change be selected beforehand. For more on

this (see Section 8.5).

Calibration

If “Calibration ON” is selected, contact output occurs during

calibration. For the definition of Under calibration (see Section 8.2.1).

Maintenance

If “Maintenance ON” is selected, contact output occurs during

maintenance. For the definition of Under maintenance (see Section

8.2.1).

Blow back

If “Blow back ON” is selected, contact output occurs during blow

back. For the definition of During blow back (see Section 8.2.1).

High limit temperature alarm Not supported by the oxygen analyzer.
Calibration gas press. low

If “Cal. gas press. low ON” is selected, contact output occurs

(“answer-back signal to a calibration gas low pressure signal”)when

a calibration gas low pressure signal is applied to the contact input.

To do this, it is required, in Input contacts setup, that “Cal. gas press.

low” be selected beforehand. For more on this (see Section 8.5).

Process upset

If “Process upset” is selected, contact output occurs (“answer-back

signal to a process upset signal) when the process upset signal is

applied to the contact input. To do this, it is required, in Input contacts

setup, that “Process upset” be selected beforehand (see Section 8.5).

Note: To provide an alarm with an output contact, be sure to make an alarm setting.