2 blow back, 1 blow back setup, Blow back -8 10.2.1 blow back setup -8 – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 146: Blow back setup

Advertising
background image

<10. Other Functions>

10-8

IM 11M12A01-02E

10.2 Blow Back

10.2.1 Blow back Setup

This section explains the parameter settings for performing blow back.

10.2.1.1 Mode

There are three modes of blow back operation: no function, semi-automatic, and automatic.

Blow back is not performed when the mode is set to “No function”. In “Semi_Auto” mode, blow

back can be started by key operation on the display or by a contact input signal, and then

sequentially performed at a preset blow back time and hold time. In “Auto” mode, blow back

is automatically performed at preset intervals. For “Semi_Auto” or “Auto” modes, blow back is

performed. The following restrictions apply:
When “No function” is selected:
Blow back is not performed
When “Semi_Auto” is selected:
Semi-auto blow back can be performed. (Blow back does not start at Auto blow back start time.)
When “Auto” is selected:
Blow back can be performed in either “Auto” or “Semi_Auto” mode.
<Setting Procedure>

(1) From the Basic panel display, touch [Setup] key; on the Execution/Setup display which

appears, select “Maintenance”.

(2) On the Maintenance display, select “Blow back setup” and the Mode selection pull down

allows you to select between “No function”, “Semi_Auto” and “Auto” (see Fig.10.8).

Enter

No function

Semi_Auto

Auto

Mode:

Blow back setup

F10.9E.ai

Figure 10.8 Blow back setup Display

8th Edition : Jan.13,2012-00