1 parts required for piping in system 1, Parts required for piping in system 1 -2, Caution – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 62

Advertising
background image

<4. Pipomg>

4-2

IM 11M12A01-02E

8th Edition : Jan.13,2012-00

Piping in System 1 is as follows:

• Connect a stop valve to the nipple at the calibration gas inlet of the detector. Then mount a

joint for a 6 mm (O.D.) x 4 mm (I.D.) soft tube at the stop valve connection hole of the inlet

side (see Section 4.1.2). The tube is to be connected to this joint only during calibration.

• If a high temperature detector is used and no piping can be installed for the reference gas,

place piping in the exhaust hole for the sample gas on the high temperature probe adapter

so that the sample gas is carried away from the vicinity of the detector (see Section 4.1.4,

Figure 4.6).

• If a high temperature detector is used and the sample gas pressure is negative, connect

an ejector assembly to the sample gas exhaust hole of the high temperature probe adapter

(see Section 4.1.4, Figure 4.3).

• If a high temperature detector is used and the pressure of the sample gas is 0.49 kPa or

higher, it is recommended that a needle valve (throttle) be used in the sample gas exhaust

of the high temperature probe adapter (see Section 4.1.4, Figure 4.4).

CAUTION

• This is for lowering the sample gas temperature below 700 °C. If the gas temperature is high

and the pressure is also significantly high, the sample gas temperature may not fall to below

700 °C before it reaches the detector.

On the other hand, if the sample gas temperature is lowered too much, condensation may

be produced in the High Temperature Probe Adapter. During wintertime, it is recommended

that the High Temperature Probe Adapter be protected with an insulating material to avoid

condensation forming (see Section 4.1.4, Figure 4.5).

For the usage of the High Temperature Probe Adapter, refer to Section 3.2.2.

4.1.1

Parts Required for Piping in System 1

Check that the parts listed in Table 4.1 are ready.

Detector

Piping location

Parts Name

Note

General use

detector

Calibration gas inlet Stop valve

Recommended by YOKOGAWA

(L9852CB or G7016XH)

Nipple *

Rc1/4 or 1/4 NPT

General parts

Joint for tube connection

Rc1/4 (1/4NPT) for a Ø6 x

Ø4 mm soft tube

General parts

Reference gas inlet (sealed up)

(when piping is required, refer to section 4.1.3)

High temperature

detector

(0.15 m)

Calibration gas inlet Stop valve

Recommended by YOKOGAWA

(L9852CB or G7016XH)

Nipple *

Rc1/4 or 1/4 NPT

General parts

Joint for tube connection

Rc1/4 (1/4 NPT) for a Ø6 x Ø4 mm soft tube

Reference gas inlet (sealed up)

(when piping is required, refer to section 4.1.3)

Sample gas outlet

Ejector assembly *

Recommended by YOKOGAWA

(E7046EC or E7046EN)

T-shaped joint of the same

diameter *

R1/4 or 1/4 NPT

General parts

Needle valve *

Rc1/4 or 1/4 NPT

General parts

Nipple of other diameter * R1/2 to R1/4 or R1/2 to 1/4

NPT

General parts

Note: Parts with marking * are used when required.

General parts can be found on the local market.