2 high temperature detector (zr22g-015), High temperature detector (zr22g-015) -2 – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 86

Advertising
background image

<6. Components>

6-2

IM 11M12A01-02E

8th Edition : Jan.13,2012-00

6.1.2

High Temperature Detector (ZR22G-015)

Sample gas outlet

When a sample gas pressure is

negative, connect the ejector assembly.

When the sample gas is high

temperature and high pressure, and does

not fall below 700 °C, connect a pressure

control valve (e.g. a needle valve).

(Refer to Section 3.2.2.)

Separate type

High Temperature Detector (ZR22G-015)

When the temperature of the sample gas is

between 700 °C and 1400 °C, mount this

detector with a ZO21P-H probe adapter.

F6.2E.ai

Flange

Selectable from JIS, ANSI,

JPI or DIN

High Temperature Probe Adapter (ZO21P-H)

The probe is made of either SUS 310S

or silicon carbide (SiC).

Its length is 0.5, 0.6, 0.7, 0.8, 0.9, 1.0 or 1.5m.

When using an SiC probe, mount it vertically

downward.

Figure 6.2 High Temperature Detector

Advertising
This manual is related to the following products: