<5. Wiring>

5-8

IM 11M12A01-02E

8th Edition : Jan.13,2012-00

Notice when closing the cover of the detector

NOTE

• Before opening the detector cover, loosen the lock screw. If the screw is not loosened first,

the screw will damage the cover, and the terminal box will require replacement.

When opening and closing the cover, remove any sand particles or dust to avoid gouging the

thread.
• After screwing the cover in the detector body, secure it with the lock screw.

Lock screw

Detector cover

F5-8E.ai

Figure 5.7

5.3.3

Connection to Converter

To connect the wiring to the converter, proceed as follows:
(1) M4 screws are used for the terminals of the converter. Each cable should be terminated in

the corresponding size crimp-on terminals.

(2) When a rubber insulated glass braided wire is used for wiring to the detector, use a terminal

box. For wiring between the terminal box and the converter, basically use cable, e.g., PVC

insulated PVC sheathed cable.

CAUTION

The above is to prevent moisture or corrosive gas from entering the converter.

Where the ambient environment of the detector and the converter is well-maintained, it is

permissible to connect the wiring from the detector directly to the converter with protection by

conduits.

WARNING

This wiring is to carry power for the heater. Be careful to wire the correct terminals, and be

careful not to ground or short circuit terminals when wiring, as otherwise the instrument may be

damaged.