<10. Other Functions>

10-1

IM 11M12A01-02E

8th Edition : Jan.13,2012-00

10. Other Functions

10.1 Display

10.1.1 Detailed-data Display

Press the [Detailed-data] key on the Basic panel display to view the detailed operation data as

shown in Figure 10.1.

Pressing the [▼] or [▲] key, you can advance the page or go back to your desired page.

• Detailed-data display
There are ten panel displays for viewing detailed data. The following briefly describe the

operational data displayed on the Detailed-data display.

Span gas ratio: 0 . 0 %

Zero gas ratio 1 0 0 . 0 %

Response time: 0 s

Cell robustness: life> 1year

Cell temperature: 750 °C

C.J.temperature: 47 °C

Tag:

Hold

F10.1E.ai

Figure 10.1 Detailed-data Display

10.1.1.1 Span gas and Zero gas Correction Ratios

These are used to check for degradation of the sensor (cell). If the correction ratio is beyond the

limits as shown in Figure 10.2, the sensor should no longer be used.
These ratios can be found by calculating the data as shown below.

21.0

p1

0.51

81.92

ez

e1

es

0

C

A

B

Zero gas correction ratio = (B/A) x 100 (%) Correctable range: 100 ± 30%

Span gas correction ratio = (C/A) x 100 (%) Correctable range: 0 ± 18%

Cell

electromotive

force, mV

Calibration curve

before correction

Previous

zero gas data

Corrected calibration curve

(theoretical calibration curve)

Zero origin

Span origin

Span gas

concentration

Oxygen concentration (vol%O

2

)

F10.2E.ai

Figure 10.2