Manual calibration – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 104

Advertising
background image

<7. Startup>

7-16

IM 11M12A01-02E

(3) Follow the display message in Figure 7.19 to turn on span gas flow. Open the span gas flow

valve for the Flow Setting Unit by loosening the valve lock nut and slowly turning the valve

shaft counterclockwise to flow the span gas at 600 ± 60 ml/min. Use the calibration gas

flowmeter to check the flow.

Enter

Open span gas valve.
Set flow span gas to
600ml/min.

Valve opened

Cancel calibration

Manual calibration

F7.19E.ai

Enter

21.00%

0.5min./div

1.00%

21.0 %O

2

CAL. TIME

Manual calibration

F7.20E.ai

Figure 7.19 Span gas Flow Display

Figure 7.20 Span gas Trend Graph

(for Manual Calibration)

(4) If “Valve opened” is selected as in Figure 7.19 , an oxygen concentration trend graph (with

the oxygen concentration being measured) appears (see Figure 7.20). The CAL.TIME in

the bottom area of the panel flashes. Observe the trend graph and wait until the measured

value stabilizes in the vicinity of 21% on the graph. At this point, calibration has not yet been

executed yet, so even if the measured value is above or below 21%, no problem occurs.

(5) After the measured value has stabilized, press the [Enter] key to display the “Span

calibration complete” display shown in Figure 7.21. At this point, the measured value is

corrected to equal the span gas concentration setting. Close the span gas flow valve. The

valve lock nut should be tightened completely so that the span gas does not leak.

Enter

Span calibration
Close the span gas valve.

Zero calibration

End

Manual calibration

F7.21E.ai

Enter

Check the zero gas
concentration value

Next

Cancel calibration

Change value: 1.00%

Manual calibration

F7.22E.ai

Figure 7.21 Span Calibration Complete

Figure 7.22 Zero Gas Concentration

Zero Calibration

Check Display

Start Display

(in Manual Calibration)

(6) Select “Zero calibration” as in Figure 7.21 to display the zero gas concentration check

display (Manual calibration) . Check that the zero gas oxygen concentration value and the

calibration gas oxygen concentration value agree. Then select “Next” as in Figure 7.22.