2 output hold setting, 1 definition of equipment status, Output hold setting -3 8.2.1 – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 109: Definition of equipment status -3

Advertising
background image

<8. Detailed Data Setting>

8-3

IM 11M12A01-02E

8th Edition : Jan.13,2012-00

8.2

Output Hold Setting

The “output hold” functions hold an analog output signal at a preset value during the equipment’s

warm-up time or caribration or if an error arises. Outputs 1 and 2 can not be set individually. Table

8.1.2 shows the analog outputs that can be retained and the individual states.

Table 8.1.2

Equipment status

Output

hold values available

During warm-up

Under

maintenance

Under calibration

During blow back

On Error

occurrence(*1)

4 mA

20 mA

Without hold feature

Retains output from just before

occurrence

Set value (2.4 to 21.6 mA)

: The output hold functions are available.

*1: The output hold functions on error occurrence are unavailable when option code “/C2” or “/C3”(NAMER NE 43 compliant) is

specified.

8.2.1

Definition of Equipment Status

(1) During warm-up
“During warm-up” is the time required after applying power until sensor temperature stabilizes at

750 ºC and the instrument is in the measurement mode.

(2) Under Maintenance
“Under maintenance” is the time that starts when you move to the Execution/Setup display by

touching the [Setup] key on the Basic panel display and ends when you return to the Basic panel

display. It includes when you operates keys on lower level menu displays of the Execution/Setup

display

.

(3) Under Calibration (see Chapter 9, Calibration, later in this manual)
For manual Calibration, “Under calibration” is the time that starts when you move to the Manual

calibration display (Figure 7.19) from the Calibration display, lasts while you are operating keys

for performing calibration manually, and ends when you press the End key and after a preset hold

time has elapsed.
For semi-automatic calibration, “Under calibration” is the time required from entering calibration

instructions to perform a, either by using the touchpanel or by a contact input, calibration until the

hold (output stabilization) time elapses.
For automatic calibration, “Under calibration” is the time required, after performing an appropriate

calibration until the hold (output stabilization) time elapses.

(4) During Blow back (see Section 10.2, Blow Back, later in this manual)

During semi-automatic blow back:
“During semi-automatic blow back” is the time required after pressing the [Blow back start] key,

by using the touchpanel or entering a blow back start instruction by using a contact input, until the

blow back time and hold (output stabilization) time elapse.

During automatic blow back:
“During automatic blow back” is the time required after reaching the blow back start time until the

blow back time and hold (output stabilization) time elapse.

(5) On Error occurrence
This is the time at which any of Errors 1 to 4 occurs.