<8. Detailed Data Setting>

8-22

IM 11M12A01-02E

8.6.5

Setting Passwords

Unauthorized access to lower level menu displays from the Execution/Setup display can be

protected by passwords. You can set separate passwords for Calibration, Blow back, and

Maintenance and for Commissioning

Proceed as follows:

1) From the Basic panel display touch the [Setup] key, and the Execution/Setup display

appears.

2) Select “Commissioning” , and the Commissioning display appears.
3) Select “Others” then “Passwords”. The display shown in Fig. 8.22 appears.

First set the password for Calibration, Indication Check, Blow back, and Maintenance:

4) Select “Calibration, Blow back, and Maintenance”.
5) A text entry display appears. Enter password as up to 8 alphanumeric characters.
6) You can enter a password for Commissioning by the same procedure.
7) Record the passwords and look after them carefully.

Enter

Calibration,Blow back

and Maintenance:


Commissioning:

Passwords

F8.22E.ai

Figure 8.22 Passwords Display

<Default setting>
The passwords are not set as shipped from factory. If you reset initialize the parameters, and

password settings are deleted.

If you forget a password, select “Commissioning” in the Execution/Setup display, and enter

“MOON.” By doing so, you can enter the Commissioning display only. Then display the

Passwords and verify the set passwords.

8th Edition : Jan.13,2012-00