3 countermeasures when measured value shows error, 1 measured value higher than true value – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 181

Advertising
background image

<12. Troubleshooting>

12-11

IM 11M12A01-02E

12.3 Countermeasures When Measured Value

Shows Error

The causes that the measured value shows an abnormal value is not always due to instrument

failures. There are rather many cases where the causes are those that measuring gas itself

is in abnormal state or external causes exist, which disturb the instrument operation. In this

section, causes of and measures against the cases where measured values show the following

phenomena will be described.
(1) The measured value is higher than the true value.
(2) The measured value is lower than the true value.
(3) The measured value sometimes shows abnormal values.

12.3.1 Measured Value Higher Than True Value

<Causes and Countermeasures>
(1) The measuring gas pressure becomes higher.

The measured oxygen concentration value X (vol% O

2

) is expressed as shown below,

when the measuring gas pressure is higher than that in calibration by ∆p (kPa).

X=Y [ 1+ (∆p/101.30) ]

where Y: Measured oxygen concentration value at the same pressure as in calibration

(vol% O

2

).

Where an increment of the measured value by pressure change cannot be neglected,

measures must be taken.

Investigate the following points to perform improvement available in each process.

• Is improvement in facility’s aspect available so that pressure change does not occur?

• Is performing calibration available under the average measuring gas pressure

(internal pressure of a furnace)?

(2) Moisture content in a reference gas changes (increases) greatly.

If air at the detector installation site is used for the reference gas, large change of moisture

in the air may cause an error in measured oxygen concentration value (vol% O

2

).

When this error is not ignored, use a gas in which moisture content is constant such as

instrument air in almost dry condition as a reference gas.

In addition, change of moisture content in exhaust gas after combustion is also considered

as a cause of error. However, normally this error is negligible.

(3) Calibration gas (span gas) is mixing into the detector due to leakage.

If the span gas is mixing into the detector due to leakage as a result of failure of the valve

provided in the calibration gas tubing system, the measured value shows a value a little

higher than normal.

Check valves (needle valves, check valves, solenoid valves for automatic calibration,

etc.) in the calibration gas tubing system for leakage. For manual valves, check them after

confirming that they are in fully closed states. In addition, check the tubing joints for leakage.

(4) The reference gas is mixing into the measuring gas and vice versa.

Since the difference between oxygen partial pressures on the sensor anode and cathode

sides becomes smaller, the measured value shows a higher value.

An error which does not appear as the Error1 may occur in the sensor. Sample gas and/or

the reference gas may be leaking. Visually inspect the sensor. If any crack is found, replace

the sensor assembly with a new one.

NOTE

Data such as cell robustness displayed in the Detailed-data display should also be used for

deciding sensor quality as references.