<12. Troubleshooting>

12-2

IM 11M12A01-02E

8th Edition : Jan.13,2012-00

12.1.2 Remedies When an Error Occurs

12.1.2.1 Error1: Cell Voltage Failure

Error1 occurs when the cell (sensor) voltage input to the converter falls below -50 mV

(corresponding to about 200 vol% O

2

). The following are considered to be the causes for the cell

voltage falling below -50 mV:
(1) Poor contact in terminal connections between the converter and detector
(2) Breakage in wiring cable between the converter and the detector
(3) Damage or deterioration of the sensor assembly
(4) Continuity failure between the sensor assembly electrode and the contact
(5) Wiring failure inside the detector
(6) Abnormality in electrical circuits in the converter

<Locating the failure and countermeasures>

No

Yes

Yes

Yes

Yes

Yes

Yes

No

No

No

No

No

F12.1.2E.ai

Turn off power to the converter.

Is there any breakage

or poor contact in the wiring to the

converter terminals?

Is there any breakage

or poor contact in the wiring to the

detector terminals?

Is the sensor extremely

dirty, corroded or broken?

Is there any

disconnection or poor continuity

in the wiring between the

detector and

converter?

Is an error indicated

on the display?

End. Carry out calibration.

Replace the damaged

portions.

Examine the wiring connection to converter terminals

"CELL+" and "CELL-" Also, check the wiring

connection in the repeater terminal box if it is used.

Examine the wiring connection status to

detector terminals 1 and 2. Also, check that

terminals and cable conductors are not corroded.

Replace the damaged

portions.

Remove the sensor assembly from the detector

and check for the presence of corrosion that may

cause a poor contact between the electrode and

the contact; also, check for dirt. A sensor assembly

in which no abnormality is found may be used again.

However, be sure to use a new metal O-ring and

contact even in such a case.

Replace the sensor

assembly.

See Section 11.1.2 for the

replacement procedure.

Remove the wiring conductors from detector

terminals 1 and 2 and short out the removed

conductors. Measure the resistance of these

wiring conductors on the converter side.

The resistance value is normal if it indicates

10 Ω or less.

Replace the wiring cable.

Turn on the power to the converter and temporarily

place the analyzer in the operating status.

Replace the sensor

assembly and temporarily

place the analyzer in the

operating status.

See Section 11.1.2 for the replacement

procedure.

The analyzer operates normally.

A failure in the detector or

the converter is suspected.

Contact Yokogawa.