<7. Startup>

7-7

IM 11M12A01-02E

7.5

Confirmation of Converter Type Setting

This converter can be used for both the Oxygen Analyzer and the Humidity Analyzer.

Before setting the operating data, be sure to check that the desired converter model has been

set.
Note that if the converter type setting is changed, the operating data that have been set are then

initialized and the default settings remain. To set the desired operating data, follow these steps:
(1) Press the [Setup] key.
(2) Use the [▼] key to select “Commissioning” and press the [Enter] key.
(3) In the password display, enter the [Enter] key. If the password is to be set again, enter the

new password (for details, see Section 8.6.5, Setting Passwords, later in this manual).

(4) The Commissioning display shown in Figure 7.6 appears. Select “Basic setup” and press

the [Enter] key.

(5) The Basic setup display shown in Figure 7.7 then appears. Confirm the currently set

converter type. If the Humidity Analyzer option /HS was selected at the time of purchase, the

converter was set for high temperature humidity use before shipment.

(6) If the converter type is to be changed, press the [Enter] key. The display shown in Figure 7.8

then appears.

(7) Use the [▼] key to select the type of equipment. Then press the [Enter] key to complete the

converter selection.

(8) If the type of converter is changed after setting the operating data, those data are then

initialized and the default settings remain. Reset the operating data to meet the new type of

equipment.

Enter

Commissioning

mA-output setup

Alarms setup

Contact setup

Others

Basic setup

F7.6E.ai

Enter

Basic setup

This instrument is for

Oxygen

Please note:changing this

Basic setting will return ALL

other setting to Default!!

Detector is ZR22

Select measure gas: Wet

F7.7E.ai

Figure 7.6 Commissioning Display

Figure 7.7 Basic Setup

This instrument is for

Please note:changing this

Basic setting will return ALL

other setting to Default!!

Detector is ZR22

Select measure gas: Wet

Enter

Basic setup

Oxygen

Humidity

F7.8E.ai

Figure 7.8 Equipment Setup