2 manual calibration, 2 manual calibration -15, Calibration – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 103: Manual calibration

Advertising
background image

<7. Startup>

7-15

IM 11M12A01-02E

(2) Span gas concentration

With “Span gas conc” selected in the Calibration setup display, display the Numeric - data

Entry display and enter an oxygen concentration value for the span gas calibration; If

instrument air is used, enter 02100 for a 21 vol%O

2

value. When using the ZO21S Standard

Gas Unit (for use of the atmospheric air as a span gas), use a portable oxygen analyzer to

measure the actual oxygen concentration, and then enter it.

NOTE

If instrument air is used for the span gas, dehumidify the air by cooling to a dew point of - 20 ºC

and remove any oil mist or dust.

7.12.2 Manual Calibration

7.12.2.1 Preliminary

Before performing manual calibration, be sure that the ZA8F Flow Setting Unit zero gas flow

setting valve is fully closed. Open the zero gas cylinder pressure reducing valve so that the

secondary pressure equals sample gas plus approx. 50 kPa (or sample gas pressure plus

approx. 150 kPa when a check valve is used, maximum pressure rating is 300 kPa).
This applies even if you are using the ZR40H Automatic Calibration Unit.

7.12.2.2 Calibration Procedures

This manual assumes that the instrument air is the same as the reference gas used for the span

gas. Follow the steps below to conduct manual calibration:
(1) Press the [Setup] key in the Basic panel display to display the Execution/Setup display.

Then select “Calibration” in the Execution/Setup display.

Enter

Calibration

Manual calibration

F7.17E.ai

Check the span gas
concentration value

Next

Cancel calibration

Change value: 21.00%

Enter

Manual calibration

F7.18E.ai

Figure 7.17 Calibration Display

Figure 7.18 Manual Calibration

(2) Press the [Enter] key to select “Span calibration”. The Manual calibration display shown

in Figure 7.18 then appears. Check that the oxygen concentration for the span gas in this

display coincides with the oxygen concentration in the calibration gas actually used. If the

check results are assumed to be OK, select “Next” in the Manual calibration display.