<10. Other Functions>

10-2

IM 11M12A01-02E

8th Edition : Jan.13,2012-00

10.1.1.2 Cell Response Time

The cell’s response time is obtained in the procedure shown in Figure 10.3. If only either a zero-

point or span calibration has been carried out, the response time will not be measured just as it

will not be measured in manual calibration.

Figure 10.3 Typical Response Time characteristic

10.1.1.3 Robustness of a Cell

The robustness of a cell is an index for predicting the remaining life of a sensor and is expressed

as one of four time periods during which the cell may still be used:
(1) more than a year
(2) more than six months
(3) more than three months
(4) less than one month
The above four time periods are tentative and only used for preventive maintenance, not for

warranty of the performance.
This cell’s robustness can be found by a total evaluation of data involving the response time, the

cell’s internal resistance, and calibration factor. However, if a zero or span calibration was not

made, the response time cannot be measured. In such a case, the response time is not used as

a factor in evaluating the cell’s robustness.

10.1.1.4 Cell Temperature

This displays the cell (sensor) temperature, which is determined from the thermocouple emf and

cold junction temperature. Normally it is 750 ºC.

10.1.1.5 C. J. Temperature

This indicates the detector terminal box temperature, which compensates for the cold junction

temperature for a thermocouple measuring the cell temperature. When the ZR22 Detector is

used, the maximum C. J. temperature will be 150 °C. If the terminal box temperature exceeds

this, take measures to shield the terminal box from heat radiation.
The maximum C. J. temperature varies depending on the type of detector.

mA

90%

100%

The response time is obtained after the corrected calibration curve has been found. The response time

is calculated, starting at the point corresponding to 10% of the analog output up to the point at 90% of

the analog output span. That is, this response time is a 10 to 90% response.

10% of analog

output span

Five minutes maximum

Response time

Start calibration

Time

Calibration

complete

F10.3E.ai