Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 32

Advertising
background image

<2. Specifications>

2-14

IM 11M12A01-02E

8th Edition : Jan.13,2012-00

Zero calibration gas concentration setting range: 0.3 to 100 vol% O

2

(in increments of

0.01 vol% O

2

in smallest units).

Span calibration gas concentration setting range: 4.5 to 100 vol% O

2

(in increments of

0.01 vol% O

2

in smallest units).

Use nitrogen-balanced mixed gas containing 0 to 10 vol% O

2

scale of

oxygen, and 80 to 100 vol% O

2

scale of oxygen for standard zero gas

and standard span gas respectively.

Calibration interval; date/time setting: maximum 255 days

l

Model and Codes

Model

Suffix code

Option

code

Description

ZR402G

- - - - - - - - - - - - - - - -

- - - - - -

Separate type Zirconia Oxygen Analyzer, Converter

Converter

thread

-P

-G

-M

-T

- - - - - -

- - - - - -

- - - - - -

- - - - - -

G1/2

Pg13.5

M20x1.5 mm

1/2NPT

Display

-J

-E

-G

-F

-C

- - - - - -

- - - - - -

- - - - - -
- - - - - -

- - - - - -

Japanese

English

German

French

Chinese

Instruction manual

-J

-E

-C

- - - - - -

- - - - - -
- - - - - -

Japanese

English

Chinese

-A

- - - - - -

Always -A

Options

Tag plates

NAMUR NE43

compliant

/H

Hood (*2)

/SCT

/PT

Stainless steel tag plate (*1)

Printed tag plate (*1)

/C2

/C3

Failure alarm down-scale:

Output status at CPU failure and hardware error is 3.6 mA or

less (*3)

Failure alarm up-scale:

Output status at CPU failure and hardware error is 21.0 mA or

more (*3)

*1 Specify either /SCT or /PT option code.

*2 Sun shield hood is still effective even if scratched.

*3 Output signal limits: 3.8 to 20.5 mA. Specify either /C2 or /C3 option code.

(Note) If AC line voltage is 125 V AC or greater, or in the EEC, the ZO21D cannot be used with the ZR402G.