<2. Specifications>

2-8

IM 11M12A01-02E

8th Edition : Jan.13,2012-00

2.3

Separate type Detector for High

Temperature and Related Equipment

2.3.1

ZR22G (0.15m) Separate type Detector for High

Temperature

Standard Specifications
Construction:

Water-resistant, non-explosionproof

Probe length:

0.15 m

Terminal box:

Aluminum alloy

Probe material:

Probe material in contact with gas: SUS 316 (JIS) (Probe), SUS 304

(JIS) (Flange), Zirconia (Sensor), Hastelloy B, (Inconel 600, 601)

Weight:

Approx. 3 kg

Installation:

Flange mounting (The use of high temperature detector probe adapter

ZO21P is necessary.)

Flange standard:

JIS 5 K 32 FF equivalent (thickness varies)

Mounting angle:

Any angle between horizontal and vertical (high temperature probe is

fitted with an adapter)

Reference gas and calibration gas piping connection: Rc 1/4 or 1/4 NPT female
Cable inlet:

G 1/2, Pg 13.5, M20 x 15, 1/2 NPT

Ambient temperature: -20 to 150 °C
Sample gas temperature: 0 to 700 °C (temperature at the measuring point of the sampling gas.

0 to 750 °C or 0 to 1400 °C when the probe adapter for high temperature

is used.

Temperature of the probe adapter shall not exceed 300 °C to protect the

gasket and avoid the bolts seizing together.

Sample gas pressure: -0.5 to +5 kPa: when used at the range of more than 0 to 25 vol% O

2

,

-0.5 to +0.5 kPa. (An ejector assembly is required for negative pressure

application.)

Model and Code:

Refer to “Model and Codes” in page 2-5.