<8. Detailed Data Setting>

8-15

IM 11M12A01-02E

8.6

Other Settings

8.6.1

Setting the Date-and-Time

The following describe how to set the date-and-time. Automatic calibration or blow back works

following this setting.

Proceed as follows:

1) From the Basic panel display if you touch the [Setup] key, the Execution/Setup display

appears.

2) Select “Commissioning” and the Commissioning display appears.
3) Select “Others” and the display of Fig 8.12 appears.
4) Select “Clock” and the display of Fig 8.13 appears.
5) Select “Set date” and the numerical entry display appears. To enter June 21. 2004 , enter

210604. Touch the [Enter] key and you revert to the display of Fig 8.13.

6) Select “Set time”. Enter the time in 24-hour format. To set 2:30 pm, enter 1430 on the

numerical display and touch enter. Touch the [Enter] key. The clock starts from 00 seconds..

Clock

Averaging

Fuel setup

Purging

Unit setup

Passwords

Defaults

Enter

Others

F8.12E.ai

Enter

Set date: 1 3 / 0 7 / 0 0

Set time: 1 4 : 3 9

1 3 J ul. 2 0 0 0
1 4 : 3 9 : 5 5

Clock

F8.13E.ai

Figure 8.12 Other Settings

Figure 8.13 Clock Display

8.6.2

Setting Periods over which Average Values Are

Calculated and Periods over which Maximum and

Minimum Values Are Monitored

The equipment enables the display of oxygen concentration average values and maximum and

minimum values under measurement (see Section 10.1.1, later in this manual).

The following section describes how to set the periods over which oxygen concentration average

values are calculated and maximum and minimum values are monitored.

8.6.2.1 Procedure

1) From the Basic panel display touch [Setup] key, and the Execution/Setup display appears.
2) Select “Commissioning” and the Commissioning display appears.
3) Select “Others” then Averaging. The display of Fig. 8.14 appears.
4) Select “Set period over which average is calculated” and numerical entry display appears.

To set 3 hours, enter 003. Entry range is 1 to 255 hours.

5) Select “Set period over which maximum and minimum is stored” and numerical entry display

appears. To set 48 hours, enter 048. Entry range is 1 to 255 hours.

8th Edition : Jan.13,2012-00