<8. Detailed Data Setting>

8-13

IM 11M12A01-02E

8.5

Input Contact Settings

8.5.1

Input Contact Functions

The converter input contacts execute set functions by accepting a remote dry-contact (“voltage-

free contact”) signal. Table 8.7 shows the functions executed by a remote contact signal.

Table 8.7 Input Contact Functions

tem

Function

Calibration gas pressure low

Contact input disables Semi-automatic or Automatic Calibration.

Measuring range change

While Contact signal is ON, range of mA-Output 1 is switched to 0-25

vol%O

2

and “Range” is displayed on the screen. See Figure 8.9.

Calibration start

Contact input starts Semi-Automatic Calibration. Calibration Mode

setting must be [Semi_Auto] or [Auto]. Contact signal must be

applied for at least 1 sec. Even if input signal continues to be applied,

calibration is not repeated unless contact input is released then

reapplied.

Process upset

(Combustible gas detection)

If the Contact signal is ON, heater power will be switched off. (A

one- to 11-second time interval single-output signal is available as

a contact signal.) If this operation starts, the sensor temperature

decreases and an error occurs. To restore it to normal, turn the power

off and then back on, or reset the analyzer.

Blow back start

Contact input starts Blow back. Contact signal must be applied for at

least 1 sec. Even if input signal continues to be applied, Blow back is

not repeated unless contact input is released then reapplied. (Refer

to Section 10.2, Blow Back.)

Tag:

21.0

%O

2

7.35mA -Output1

7.35mA -Output2

Range

F8.9E.ai

Figure 8.9 Changing Measuring Range with Input Contact

CAUTION

1. Measurement range switching function by an external contact input is available for analog

output1 only.

2. When making a semi-automatic calibration, be sure to set the semi-automatic or automatic

mode using the Calibration setup display.

3. When carrying out “Blow back,” be sure to set “Blow back” in the output contact setup.
4. When the combustible gas detection signal is sent to the contact input, the converter will cut

the power supply to the heater of the detector. As a result, the heater temperature becomes

low and Error 1 or Error 2 happens.