Startup, 1 checking piping and wiring connections, 2 checking valve setup – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 89: 3 supplying power to the converter, Startup -1, Checking piping and wiring connections -1, Checking valve setup -1, Supplying power to the converter -1

Advertising
background image

<7. Startup>

7-1

IM 11M12A01-02E

8th Edition : Jan.13,2012-00

7.

Startup

The following describes the minimum operating requirements — from supplying power to the

converter to analog output confirmation to manual calibration.
System tuning by the HART communicator, refer to IM11M12A01-51E “ HART Communication

Protocol ”.

7.1

Checking Piping and Wiring Connections

Check that the piping and wiring connections have been properly completed in accordance with

Chapter 4, “Piping,” and Chapter 5, “Wiring.”

7.2

Checking Valve Setup

Set up valves and associated components used in the analyzer system as follows:
(1) If a stop valve is used in the detector’s calibration gas inlet, fully close this valve.
(2) If instrument air is used as the reference gas, adjust the air-set secondary pressure so that

an air pressure equals sample gas pressure plus approx. 50 kPa (or sample gas pressure

plus approx. 150 kPa when a check valve is used, maximum pressure rating is 300 kPa) is

obtained. Turn the reference gas flow setting valve in the flow setting unit to obtain a flow

of 800 to 1000 ml/min. (Turning the valve shaft counterclockwise increases the rate of flow.

Before turning the valve shaft, if the valve has a lock nut, first loosen the lock nut.) After

completing the valve setup, be sure to tighten the lock nut.

NOTE

The calibration gas flow setting is described later. Fully close the needle valve in the flow setting

unit.

7.3

Supplying Power to the Converter

CAUTION

To avoid temperature changes around the sensor, it is recommended that (rather than turning it

on and off) power be continuously supplied to the Oxygen Analyzer if it is used in an application

where it is used periodically.
It is also recommended to flow a span gas (instrument air) beforehand.

Supply power to the converter. A display as in Figure 7.1, which indicates the detector’s sensor

temperature, then appears. As the heat in the sensor increases, the temperature gradually rises

to 750

°C. This takes about 20 minutes after the power is turned on, depending somewhat on the

ambient temperature and the sample gas temperature.
After the sensor temperature has stabilized at 750

°C, the converter is in measurement mode.

The display panel then displays the oxygen concentration as in Figure 7.2. This is called the

Basic panel display.

Advertising
This manual is related to the following products: