2 remedies when alarms are generated, 2 remedies when alarms are generated -6 – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 176

Advertising
background image

<12. Troubleshooting>

12-6

IM 11M12A01-02E

12.2.2 Remedies When Alarms are Generated

12.2.2.1 Alarm 1: Oxygen Concentration Alarm

This alarm is generated when a measured value exceeds an alarm set point or falls below it.

For details on the oxygen concentration alarm, see Section 8.3, Setting Oxygen Concentration

Alarms, in the chapter on operation.

12.2.2.2 Alarm 6: Zero-point Calibration Coefficient Alarm

In automatic or semi-automatic calibration, this alarm is generated when the zero correction ratio

is out of the range of 100 ± 30% (refer to Section 9.1.3, Compensation).

The following are possible causes of this alarm.
(1) The zero-gas oxygen concentration does not agree with the value of the zero-gas

concentration set in Calibration setup. Otherwise, the span gas is used as the zero gas.

(2) The flow rate of zero gas is out of the specified flow (600 ± 60 ml/min).
(3) The sensor assembly is damaged and so cell voltage is not normal.

<Locating cause of failure, and countermeasures>
(1) Check that the following have been set up correctly. If not, correct them. Then, recalibrate.

a. Check the preset zero gas concentration on the Calibration setup display. The displayed

concentration value has agreed with the concentration of the zero gas actually used.

b. The piping for calibration gases has been constructed so that the zero gas does not leak.

(2) If the alarm is not generated during the recalibration, improper calibration conditions are

considered as the cause of the alarm generated in the previous calibration. In this case, no

particular restoration is necessary.

(3) If the alarm is generated again during the recalibration, deterioration of or damage to the

sensor assembly is considered as the cause of the alarm. It is necessary to replace the cell

with a new one. Before replacement, carry out the following.

Check the cell voltages when the zero gas and span gas are introduced.

a. From the Basic panel display, move to the Detailed-data display of the channel where the

alarm is being generated.

b. When the [▼] key is pressed once, the cell voltage should be indicated on the top line

(Figure 12.5).

c. Check whether or not the displayed cell voltage is very different from the theoretical value

at the oxygen concentration of the relevant gas. See Table 12.3 for the theoretical cell

voltages. Although the tolerance to the theoretical value cannot be generally specified, a

reasonable one may be approximately ±10 mV.

Table 12.3 Oxygen Concentration and Cell Voltage

Oxygen concentration

(vol%O2)

Cell voltage

(mV)

1%

67.1

21%

0

8th Edition : Jan.13,2012-00

Advertising
This manual is related to the following products: