<4. Pipomg>

4-8

IM 11M12A01-02E

8th Edition : Jan.13,2012-00

4.3

Piping for System 3

Piping in System 3 is illustrated in Figure 4.10. In System 3, calibration is automated;

however, the piping is basically the same as that of System 2. Refer to Section 4.2.
Adjust secondary pressure of both the air set and the zero gas reducing valve so that these

two pressures are approximately the same. The flow rate of zero and span gases (normally

instrument air) are set by a single needle valve.
After installation and wiring, check the calibration contact output (see Sec. 7.11.2), and adjust

zero gas reducing valve and calibration gas needle valve so that zero gas flow is within the

permitted range. Next check span gas calibration contact output and adjust air set so that span

gas flow is within the permitted range.

CHECK

OUT

ZERO GAS IN

AIR IN

Instrument air Approx.

1.5 l/min.

Flowmeter

Flowmeter

REF

OUT

*2

*2

EV1

EV2

Solenoid valve

EV1, EV2

F2-6-3E.ai

*2: Needle valve is supplied as accessory with flowmeter

Figure 4.9 Piping inside in System 3

When blow back function is used by contact input to the ZR402G converter, install blow back

piping as illustrated in Section 4.3.1.

NOTE

Blow back function means the function to get rid of dust inside a probe in a high temperature

probe adapter by using compressed air, when a high temperature detector is used.

Instrument air

Pressure

reducing

valve

Air Set

Needle

valve

Flowmeter

Reference gas

Calibration gas

Check valve

~

Zero gas cylinder

ZR40H

Automatic Calibration Unit

ZR22G Separate type

Zirconia Oxygen Analyzer, Detector

ZR402G Converter

Calibration gas

unit case

F4-9E.ai

EXA ZR402G

100 to 240 V AC

Contact input

Analog output, contact output

Digital output (HART)

Signal

(6-core shield cable)

Heater (2-core cable)

Figure 4.10 Piping for System 3