<8. Detailed Data Setting>

8-21

IM 11M12A01-02E

8.6.4

Setting Purging

Purging is to remove condensed water in the calibration gas pipe by supplying a span calibration

gas for a given length of time before warm-up of the detector. This prevents cell breakage during

calibration due to condensed water in the pipe.
Open the solenoid valve for the automatic calibration span gas during purging and after the purge

time has elapsed, close the valve to start warm-up.
Purging is enabled when the cell temperature is 100 °C or below upon power up and the purge

time is set in the range of 1 to 60 minutes.

Tag:

30

°C

4mA -Output1

4mA -Output2

Hold

Purge

F8.19.ai

Figure 8.19 Display during Purging

8.6.4.1 Procedure

Set the purging time as follows:
(1) Press the [Setup] key in the Basic panel display to display the Execution/Setup display.
(2) Select “Commissioning” in the Execution/Setup display. The Commissioning (Setup) display

then appears.

(3) Select “Others” in that display and the Others display then appears, as shown in Figure

8.20.

(4) Select “Purging”. The purging time setting display appears, as shown in Figure 8.21.
(5) Point to the “Purging time” and press the [Enter] key. Then the display for selecting purging

time appears.

(6) Enter the desired numeric value from the numeric-data entry display.

The allowable input ranges from 0 to 60 minutes.

Enter

Others

F8.20.ai

Clock

Averaging

Fuel setup

Purging

Unit setup

Passwords

Defaults

Enter

Purging time: 0 min

Purging

F8.21E.ai

Figure 8.20 Other Settings

Figure 8.21 Purging Display

8th Edition : Jan.13,2012-00