Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 64

Advertising
background image

<4. Pipomg>

4-4

IM 11M12A01-02E

8th Edition : Jan.13,2012-00

If the temperature of the sample gas exceeds the specified value and its pressure exceeds 0.49 k

Pa, the sample gas temperature may not be below 700 °C at the detector.

In such a case, connect a needle valve (found on the local market) through a nipple (also found

on the local market) to the probe adapter sample gas exhaust (Rc 1/2) so that the sample gas

exhaust volume is restricted.

Reducing nipple

(R1/2-R1/4 or R1/2-1/4NTP)

Needle valve

F4-4E.ai

Figure 4.4 Mounting the needle valve for restricting exhaust flow of the sample gas

In cases where condensation is likely to occur in the probe adapter when the sample gas is

cooled, protect the probe adapter with an insulating material as illustrated in Figure 4.5.

Sample gas outlet

Cover flange

Adapter for high temperature probe

Detector

Insulating material

Probe

F4-5E.ai

Figure 4.5 Preventing to condensation

If the sample gas is to be vented at a distance from the detector because no reference gas piping

can be provided, an exhaust pipe should be installed as illustrated in Figure 4.6.

In addition, the exhaust pipe shall be kept warm to protect against condensation.

Exhaust pipe

Rc1/2

Use thick piping for exhaust.

Horizontal part should be kept to the

minimum to maintain piping strength.

F4-6E.ai

Figure 4.6 Exhaust pipe