2 displays and remedies when alarms are generated, 1 alarm types – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 175

Advertising
background image

<12. Troubleshooting>

12-5

IM 11M12A01-02E

12.2 Displays and Remedies When Alarms are

Generated

12.2.1 Alarm Types

When an alarm is generated, the alarm indication blinks in the display to notify of the alarm

(Figure 12.3). Pressing the alarm indication displays a description of the alarm.

Alarms include those shown in Table 12.2.

Tag:

21.0

%O

2

17.42mA -Output1

17.42mA -Output2

F12.3E.ai

Tag:

Alarm 6 :Zero conc.ratio

F12.4E.ai

Figure 12.3

Figure 12.4

Table 12.2 Types of Alarms and Reasons for Occurrence
Alarm

Code

Alarm Type

Occurrence Conditions

Alarm 1 Oxygen concentration alarm

Measured oxygen concentration value exceeds or falls below

the preset alarm limits. (refer to Section 8.3, Setting Oxygen

Concentration Alarms).

Alarm 6 Zero-point calibration coefficient

alarm

In automatic or semi-automatic calibration, zero correction

ratio is outside the range of 100 ±30%. (refer to Section 9.1.3,

Compensation).

Alarm 7 Span-point calibration coefficient

alarm

Span correction ratio is outside the range of 0±18% (refer to

Section 9.1.3, Compensation).

Alarm 8 EMF stabilization time-up alarm

In automatic or semi-automatic calibration, cell voltage does not

stabilize after calibration time is up.

Alarm 10 Cold junction temperature alarm

Temperature of the cold junction placed in the detector terminal

box falls below -25 ºC or exceeds 155 ºC.

Alarm 11 Thermocouple voltage alarm

Generated when thermocouple voltage exceeds 42.1 mV (about

1020 ºC) or falls below -5 mV (about -170 ºC).

Alarm 13 Battery low alarm

Internal battery needs replacement.

If an alarm is generated, actions such as turning off the heater power are not carried out. The

alarm is cancelled when the cause of the alarm is removed. However, Alarm 10 and/or Alarm 11

may be generated concurrently with Error2 (heater temperature error). In this case, the operation

when the error occurs has priority.
If the power to the converter is turned off after an alarm is generated and the converter is restarted

before the cause of the alarm has been removed, the alarm will be generated again. However,

Alarms 6, 7, and 8 (alarms related to calibration) are not generated unless calibration is executed.

8th Edition : Jan.13,2012-00

Advertising
This manual is related to the following products: