<12. Troubleshooting>

12-9

IM 11M12A01-02E

(3) A failure of the cold junction temperature sensor located at the detector terminal block

occurred.

(4) A failure of the electrical circuits inside the converter occurred.

If “C.J.temperature” exceeds 150 °C or falls below -20 °C, the following can be considered.

• The temperature of the detector terminal block is out of the operating temperature range

(-20 °C to 150 °C).

• A failure of the cold junction temperature sensor located at the detector terminal block

occurred.

• A failure of the electrical circuits inside the converter occurred.

<Locating cause of failure and countermeasures>
Before proceeding to the following troubleshooting procedure, examine whether or not the

temperature of the detector terminal block is out of the operating temperature range.

The operating temperature range varies with the type of detector. If the detector terminal block

is out of its operating temperature range, take the measure to lower the temperature, such as

situating it so that it is not subjected to radiant heat.

The case where the Model ZR22 Detector is used:
(1) Stop the power to the converter.
(2) Remove the wiring from terminals 5 and 6 of the detector and measure the resistance

between these terminals. If the resistance value is out of the range of 1 to 1.6 kΩ, the cold

junction temperature sensor is considered to be faulty. Replace that temperature sensor

with a new one.

(3) If the resistance value is within the above range, the cold junction temperature sensor

seems to be normal. Check whether or not the cable is broken or shorted out, and whether

the cable is securely connected to the terminals. Also, check that the resistance of the wiring

between the converter and detector is 10 Ω or less.

(4) If there is no failure in the wiring, the electrical circuits inside the converter may possibly fail.

Contact the service personnel at Yokogawa.

The case where the Model ZO21D Detector is used:
(1) Without stopping the power to the converter, remove the wiring from terminals 5 and 6 of

the detector and measure the voltage between these terminals. If the voltage between the

terminals is out of the range of 0.4 to 0.7 V, the cold junction temperature sensor seems to

be faulty. Replace the cold junction temperature sensor.

(2) If the voltage between the terminals is within the above range, the cold junction temperature

sensor seems to be normal. Check whether or not the cable is broken or shorted out, and

whether the cable is securely connected to the terminals. Also, check that the resistance of

the wiring between the converter and detector is 10 Ω or less.

(3) If there is no failure in the wiring, the electrical circuits inside the converter may possibly fail.

Contact the service personnel at Yokogawa.

CAUTION

• The operating temperature range of the Model ZO21D Detector is -10 °C to 80 °C (except

for the high temperature detector ZO21D-H). Since a cold junction temperature alarm for

this analyzer is not generated until the temperature exceeds 155 °C, if the Model ZO21D

Detector is used, be careful in controlling the ambient temperature of the terminal block.

8th Edition : Jan.13,2012-00