3 setting fuels, Setting fuels -16, Averaging – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 122

Advertising
background image

<8. Detailed Data Setting>

8-16

IM 11M12A01-02E

Set period over which

average is calculated:

1 h

Set period over which

maximum and minimum is

stored: 2 4 h

Enter

Averaging

F8.14E.ai

Figure 8.14 Setting Average-Value Calculation Periods and

Maximum- and Minimum-Value Monitoring Periods

8.6.2.2 Default Values

“Set period over which average is calculated” is set to 1hr, and “Set period over which maximum

and minimum is stored” is set to 24 hrs prior factory shipment or after data initialization.. .

8.6.3

Setting Fuels

8.6.3.1 Input Parameters

The analyzer calculates the moisture content contained in exhaust gases. The following sets

forth the fuel parameters necessary for calculation and their entries. The moisture quantity may

be mathematically expressed by:

(water vapor caused by combustion and water vapor contained in the exhaust gas)

Moisture quantity =

= (Gw + Gw1)/G x 100

=

where,

Ao : Theoretical amount of air per unit quantity of fuel, m

3

/kg (or m

3

/m

3

) ... 2 in Table 8.8

G: Actual amount of exhaust gas (including water vapor) per unit quantity of fuel, m

3

/kg (or m

3

/m

3

)

Gw : Water vapor contained in exhaust gas per unit quantity of fuel (by hydrogen

and moisture content in fuel), m

3

/kg (or m

3

/m

3

)

............ 1 in Table 8.8

Gw1: Water vapor contained in exhaust gas per unit quantity of fuel (moisture content

in air), m

3

/kg (or m

3

/m

3

)

Go: Theoretical amount of dry exhaust gas per unit quantity of fuel, m

3

/kg (or m

3

/m

3

)

m: Air ratio

X : Fuel coefficient determined depending on low calorific power of fuel, m

3

/kg (or m

3

/m

3

) ....... 3 in Table 8.8

Z : Absolute humidity of the atmosphere, kg/kg ....... Figure 8.16

X + Ao x m

Gw + (1.61 x Z x m x Ao )

actual exhaust gas (including water vapor) per fuel

+ (water vapor contained in air for combustion)

x 100

Gw + (1.61 x Z x m x Ao)

Go + Gw + [(m - 1) Ao + (1.61 x Z x m x Ao)]

x 100

x 100

..... Equation 1

............ Equation 2

Fill in the boxes with fuel parameters in Equation 2 above to calculate the moisture content.

Use Ao, Gw and X shown in Table 8.8. If there are no appropriate fuel data in Table 8.8, use the

following equations for calculation. Find the value of “Z” in Equations 1 and 2 using Japanese

Standard JIS B 8222. If a precise measurement is not required, obtain the value of “Z” using a

graph for the absolute humidity indicated by a dry and wet bulb hygrometer.

8th Edition : Jan.13,2012-00