Caution – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 180

Advertising
background image

<12. Troubleshooting>

12-10

IM 11M12A01-02E

12.2.2.6 Alarm 11: Thermocouple Voltage Alarm

This alarm is generated when the emf (voltage) of the thermocouple falls below -5 mV (about

-170 °C) or exceeds 42.1 mV (about 1020 °C). Whenever Alarm 11 is generated, Error2 (heater

temperature failure) occurs.
(1) Breakage of the heater thermocouple signal wire between the converter and the detector

occurred, or the cable is not securely connected to the connecting terminals.

(2) The positive and negative poles of the heater thermocouple signal wiring are shorted out in

the wiring extension or at the connection terminals.

(3) A failure of the thermocouple at the detector heater assembly occurred.
(4) A failure of the electrical circuits inside the converter occurred.

<Locating cause of failure, and countermeasures>
(1) Stop the power to the converter.
(2) Remove the wiring from terminals 3 and 4 of the detector and measure the resistance

between these terminals. If the resistance value is 5 Ω or less, the thermocouple seems to

be normal. If it is higher than 5 Ω, it may indicate the possibility that the thermocouple has

broken or is about to break. In this case, replace the heater unit (refer to Section 11.1.3,

Replacement of the Heater Unit).

CAUTION

• Measure the thermocouple resistance value after the difference between the detector tip

temperature and ambient temperature falls to 50 °C or less. If the thermocouple voltage is

large, accurate measurement cannot be achieved.

(3) If the thermocouple is normal, check whether or not the wiring cable is broken or shorted

out, and also whether the wiring cable is securely connected to the terminals.

Also check that the wiring resistance between the converter and the detector is 10 Ω or less.

(4) If there is no failure in the wiring, the electrical circuits inside the converter may possibly fail.

Contact the service personnel at Yokogawa.

12.2.2.7 Alarm 13: Battery Low Alarm

An internal battery is used as backup for the clock. After this alarm occurs, removing power from

the instrument may cause the clock to stop but should not affect stored parameters. The internal

clock is used for blow back scheduling; if you use this then after a battery alarm occurs (until the

battery is replaced) be sure to check / correct the date and time every time you turn on the power.

<Corrective action>
When the battery low alarm occurs, remember that the battery cannot be replaced by the user.

Contact your Yokogawa service representative.

NOTE

Battery life varies with environmental conditions.
* If power is applied to the instrument continuously, then the battery should not run down, and

life is typically about ten years. However the battery will be used during the time interval

between shipment from the factory and installation.

* If power is not applied to the instrument, at normal room temperatures of 20 to 25 °C then

battery life is typically 5 years, and outside this range but within the range -30 to +70 °C then

battery life is typically 1 year.

8th Edition : Jan.13,2012-00

Advertising
This manual is related to the following products: