<9. Calibration>

9-9

IM 11M12A01-02E

9.2.3

Calibration

NOTE

1) Perform calibration under normal working conditions (e.g. continuous operation with sensor

mounted on furnace).

2) Perform both Span and Zero calibration for best resultant accuracy.
3) When instrument air is used for the span calibration, remove the moisture from the

instrument air at a dew-point temperature of -20 °C and also remove any oily mist and dust

from that air.

4) If dehumidifying is not enough, or if foul air is used, the measurement accuracy will be

adversely affected.

9.2.3.1 Manual Calibration

For manual calibration, consult Section 7.12, Calibration, earlier in this manual.

9.2.3.2 Semi-automatic Calibration

To start calibration, follow these steps:
(1) Press the [Setup] key in the Basic panel display to display the Execution/Setup display.

Then select “Calibration” from the Execution/Setup display. The Calibration display shown in

Figure 9.8 appears.

(2) Select “Semi-auto calibration” to display the Semi-auto calibration display shown in Figure

9.9.

(3) Select “Start calibration”. The display shown in Figure 9.10 appears, and then start

calibration.

Enter

Manual calibration

Semi-auto calibration

Calibration

F9.8E.ai

Check settings

Zero gas conc: 1 . 0 0 %

Span gas conc: 2 1 . 0 0 %

Cal time: 1 0 min 0 0 s

Go to setup to change value.

Start calibration

Cancel calibration

Enter

Semi-auto calibration

F9.9E.ai

Figure 9.8 Calibration Display

Figure 9.9 Semi-auto Calibration Display

2 1 . 0 0%

0.5min./div

1 . 0 0%

2 1 . 0 % O2

CAL. TIME

Semi-auto calibration

F9.10E.ai

Figure 9.10 Semi-auto Calibration Display

8th Edition : Jan.13,2012-00