<4. Pipomg>

4-6

IM 11M12A01-02E

8th Edition : Jan.13,2012-00

4.2.1

Piping Parts for System 2

Check that the parts listed in Table 4.2 are ready.

Table 4.2 Piping Parts

Detector

Piping location

Parts Name

Note

General use

detector

Calibration gas inlet Stop valve or check valve Recommended by YOKOGAWA (L9852CB or

G7016XH) Provided by YOKOGAWA

(K9292DN or K9292DS)

Nipple *

Rc1/4 or 1/4 NPT

General parts

Zero gas cylinder

User´s scope

Pressure reducing valve

Recommended by YOKOGAWA (G7013XF or

G7014XF)

Joint for tube connection

Rc1/4 or 1/4 NPT

General parts

Reference gas inlet Air set

Recommended by YOKOGAWA (G7003XF/

K9473XK or G7004XF/K9473XG)

Joint for tube connection

Rc1/4 or 1/4 NPT

General parts

High temperature

Detector

(0.15 m)

Calibration gas inlet Stop valve or check valve Recommended by YOKOGAWA (L9852CB or

G7016XH) Provided by YOKOGAWA

(K9292DN or K9292DS)

Nipple *

Rc1/4 or 1/4 NPT

General parts

Zero gas cylinder

User´s scope

Pressure reducing valve

Recommended by YOKOGAWA (G7013XF or

G7014XF)

Joint for tube connection

Rc1/8 or 1/8 NPT

Reference gas inlet Air set

Recommended by YOKOGAWA (G7003XF/

K9473XK or G7004XF/K9473XG)

Joint for tube connection

Rc1/4 or 1/4 NPT

General parts

Sample gas outlet

Ejector assembly *

Recommended by YOKOGAWA (E7046EC or

E7046EN)

T-shaped joint of the same

diameter *

R1/4 or 1/4 NPT

General parts

Needle valve *

Rc1/4 or 1/4 NPT

General parts

Nipple of other diameter * R1/2 to R1/4 or R1/2 to 1/4

NPT

General parts

Note: Parts with marking * are used when required.

General parts can be found on the local market.

4.2.2

Piping for the Calibration Gas

This piping is to be installed between the zero gas cylinder and the ZA8F flow setting unit, and

between the ZA8F flow setting unit and the ZR22G detector.
The cylinder should be placed in a case assemble E7044KF or the like to avoid any direct

sunlight or radiant heat so that the gas cylinder temperature does not exceed 40

°C.

Mount a reducing valve (specified by YOKOGAWA) on the cylinder.
Mount a check valve or stop valve (specified by YOKOGAWA) on the nipple (found on the local

market) at the calibration gas inlet of the detector as illustrated in Figure 4.8.

(The check valve or the stop valve may have been mounted on the detector when shipped.)

Connect the flow setting unit and the detector to a stainless steel pipe 6 mm (O.D.) x 4 mm or

larger (I.D.) (or nominal size 1/4 inch).