<8. Detailed Data Setting>

8-14

IM 11M12A01-02E

8.5.2

Setting Procedure

This setting example shows how to set “When contact input 1 Opens, Start Semi_Auto

Calibration”.

1) From the Basic panel display touch [Setup] key, and the [Execution/Setup] display appears.
2) Select “Commissioning” and the Commissioning display appears.
3) Select “Contact Setup” then “Input contacts”.
4) Select “Input 1” . Function Selection window (Fig 8.11) appears.
5) Select “Calibration start”.
6) For Input contact 1 select (operate if) “Closed”. (You can select “Open” or “Closed”).
7) Change Closed to Open.

Input1: No function

Input1 contact is active at:

Closed

Input2: No function

Input2 contact is active at:

Closed

Enter

Input contacts

F8.10E.ai

Input1: No function

Input1 contact is active at:

Closed

Input2: No function

Input2 contact is active at:

Closed

Enter

Input contacts

No function

Cal. gas press. low

Range change

Calibration start

Process up set

Blow back start

F8.11.ai

Figure 8.10 Input Contacts Display

Figure 8.11 Input Contacts Display

8.5.3

Default Values

All contact inputs are set to “No function”(disabled) prior factory shipment or after data

initialization..