2 grounding wiring, 6 contact output wiring, 1 cable specifications – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 82: 2 wiring procedure, Grounding wiring -10, Contact output wiring -10 5.6.1, Cable specifications -10, Wiring procedure -10

Advertising
background image

<5. Wiring>

5-10

IM 11M12A01-02E

8th Edition : Jan.13,2012-00

5.5.2

Grounding Wiring

The ground wiring of the detector should be connected to the ground terminal of the detector

case. The ground wiring of the converter should be connected to either the ground terminal of the

converter case or the protective ground terminal in the equipment.

The grounding terminals of the detector and the converter are of size M4. Proceed as follows:
(1) Keep ground resistance to 100 Ω or less (equivalent JIS D style (Class 3)).
(2) When the ambient temperature of the wiring installation is 80 to 150

°C for the wiring of the

detector, use wiring material with sufficient heat resistance.

(3) When connecting the ground wiring to the ground terminal of the converter case, be sure

that the lock washer is in contact with the case surface (see Figure 5.9.).

(4) Ensure that the jumper plate is connected between the G terminal and the protective ground

terminal of the converter.

5.6

Contact Output Wiring

Contact outputs 1 to 3 can be freely assigned to “low limit alarm”, “high limit alarm”, etc. user

selectable, but the assignment of contact output 4 is fixed (“error output”). And the action (contact

closed on error output) also cannot be changed.
When using these contact outputs, install the wiring as follows:

#1 Output

#2 Output

#3 Output

#4 Output

DO-1
DO-1

DO-2
DO-2

DO-3
DO-3

Terminal box

Annunciator or the like

F5-11E.ai

ZR402G

Converter

DO-4
DO-4

Figure 5.10 Contact output wiring

5.6.1

Cable Specifications

Number of wire in cable varies depending on the number of contact used.

5.6.2

Wiring Procedure

(1) M4 screws are used for the terminals of the converter. Each cable should be terminated

corresponding to crimp-on terminals.

(2) The capacities of the contact output relay are 30 V DC 3 A, 250 V AC 3 A. Connect a load

(e.g. pilot lamp and annunciator) within these limits.

Advertising
This manual is related to the following products: