2 default values, Default values -8 – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 136

Advertising
background image

<9. Calibration>

9-8

IM 11M12A01-02E

NOTE

When setting calibration timing requirements, bear the following precautions in mind:
(1) If the calibration interval is shorter than the sum of hold (output stabilization) time plus

calibration time, the second calibration start time will conflict with the first calibration. In such

a case, the second calibration will not be conducted. (When both zero and span calibrations

are to be performed, the calibration time is double that required for a single (zero or span)

calibration.)

(2) For the same reason, if the calibration start time conflicts with manual calibration or semi-

automatic calibration, the current calibration will not be conducted.

(3) If the calibration time conflicts with maintenance service or blow back operations, calibration

will start after completing the maintenance service or blow back operations (see Section

8.2.1, earlier in this manual).

(4) If 000 days, 00 hours are set for the calibration intervals, only the first calibration will be

conducted; a second or later calibration will not be conducted.

(5) If a past date is set to the calibration start day, no calibration will be conducted.

9.2.2

Default Values

When the analyzer is delivered, or if data are initialized, the calibration settings are by default, as

shown in Table 9.1.

Table 9.1 Default Settings for Calibration

Item

Default Setting

Calibration mode

Manual

Calibration procedure

Span - zero

Zero gas (oxygen) concentration

1.00%

Span gas (oxygen) concentration

21.00%

Hold (output stabilization) time

10 minutes, 00 seconds

Calibration time

10 minutes, 00 seconds

Calibration interval

30 days, 00 hours

Start date

01 / 01 / 00

Start time

0:00