2 connection to the detector, 3 connection to the converter, Connection to the detector -6 – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 78: Connection to the converter -6

Advertising
background image

<5. Wiring>

5-6

IM 11M12A01-02E

8th Edition : Jan.13,2012-00

5.2.2

Connection to the Detector

To connect cables to the detector, proceed as follows:
(1) Mount conduits of the specified thread size or cable glands to the wiring connections of the

detector.

The detector may need to be removed in future for maintenance, so be sure to allow

sufficient cable length.

(2) If the ambient temperature at the location of wire installation is 80 to 150

°C, be sure to use a

flexible metallic wire conduit. If a non-shielded “ 600 V silicon rubber insulated glass braided

wire “ is used, keep the wire away from noise sources to avoid noise interference.

(3) Figure 5.5 shows the layout of the detector terminals.

CELL +

TC +

TC -

TC

CELL

CJ

1

2 3

4 5

6

8

7

H T R

CELL

(+)

CELL

(-)

TC

(+)

TC

(-)

CJ

(+)

CJ

(-)

HTR HTR

F5-6E.ai

To converter,

or

terminal box

To ground

To converter,

or

terminal box

Figure 5.5 Detector terminals

The sizes of the terminal screw threads are M3.5 except for the M4 on grounding terminal.

Each wire in the cable should be terminated in the corresponding size of crimp terminal (*1)

respectively.

*1: If the ambient temperature at the detector installation site exceeds 60

°C, use a “ bare crimp-on terminal”.

(4) Except when “600 V silicon rubber insulated glass braided wire” is used, connect the cable

shield to the FG terminal of the converter.

5.2.3

Connection to the Converter

To connect the wiring to the converter, proceed as follows:
(1) M4 screws are used for the terminals of the converter. Each wire in the cable should be

terminated corresponding to crimp-on terminals.

(2) When a rubber insulated glass braided wire is used for wiring to the detector, use a terminal

box. For wiring between the terminal box and the converter, basically use cable, e.g. PVC

insulated PVC sheathed cable, rather than individual wires.

NOTE

The above is to prevent moisture or corrosive gas from entering the converter and to ground the

detector without fail.