<4. Pipomg>

4-10

IM 11M12A01-02E

8th Edition : Jan.13,2012-00

4.4

Piping for the Detector with Pressure

Compensation

The ZR22G----P Detector with Pressure Compensation may be used in System 2 and

System 3. However, it cannot use piping for high temperature probe adapter or blow back piping.

Use this detector whenever the furnace pressure exceeds 5 kPa (see Note). Even if the furnace

pressure is high, the detector can measure by adjusting pressure of the detector to the furnace

pressure using instrument air. The inside pressure of the probe will be kept the same as the

furnace pressure by feeding instrument air at higher pressure than that in the furnace.

NOTE

Sample gas pressure should not vary rapidly and widely.

*1

*1

*1

*1

Flange

156

87

Rc1/4 or 1/4NPT

Calibration gas inlet

25

48

Reference gas outlet

Flange

PIPING

PIPING

Stop Valve

Rc1/4 or 1/4NPT

Reference gas inlet

2-G1/2, 2-1/2NPT etc.

Cable connection port

*1 Dimensions may vary depending on the detector type. Refer to Section 2.2

Ensure that the flange gasket does not block the reference gas outlet.

Where necessary, make a notch on the flange

F4-12E.ai

*1

Figure 4.13 Detector with Pressure Compensation

Valve operation
1. For safety, stop the furnace that the detector is to be installed in. If furnace internal pressure

is high, this is especially dangerous.

2. Before starting instrument air flow, completely shut the stop valve in front of the reference

gas outlet.

3. Check that the reference gas outlet is not blocked by a flange gasket or the like.
4. Set the instrument air pressure higher than furnace internal pressure.
5. Completely open the stop valve in front of the reference gas outlet and, after turning on

instrument air flow, start furnace operation. As furnace internal pressure rises, confirm

that instrument air continues to flow and adjust the valve or increase supply pressure if

necessary.

6. After furnace internal pressure stabilizes, adjust flow.
7. If furnace operation is stopped, stop instrument air flow and completely shut the stop valve

in front of the reference gas outlet. You may leave reference gas flowing if you wish.