Warning, Manual calibration – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 157

Advertising
background image

<10. Other Functions>

10-19

IM 11M12A01-02E

(1) Use the needle of the zero gas valve “ CHECK GAS “ to puncture a hole in the gas cylinder

installed as described in Section 10.5.2. Fully clockwise turn the valve regulator by hand.

(2) Next, adjust the flow rate to 600 ± 60 ml/min (the flow check ball stops floating on the green

line when the valve is slowly opened). Turn the regulator of the zero gas valves back slowly

counterclockwise. At that time, the flow rate also decreases as the inner pressure of the gas

cylinder decreases. Therefore, monitor the flow check and, when the ball’s position changes

greatly, readjust the valve.

(3) Select “Valve opened” (to start calibration) from the Manual calibration display. Check the

Trend graph display to see that the measured value is stabilized. Then press the [Enter] key.

The Manual calibration display shown in Figure 10.18 appears. Then stop the zero gas flow

immediately. Turn the zero gas valve regulator fully clockwise. If this valve regulator is not

properly adjusted, the needle valve will not close completely and a cylinder gas may leak.

Enter

Zero calibration
Close the zero gas valve.

Span calibration

End

Manual calibration

F10.18E.ai

Figure 10.18 Zero gas Calibration Complete (in Manual Calibration Display)

NOTE

Be sure not to terminate a calibration in progress because of a shortage of gas in the cylinder.

Each gas cylinder is operable for nine minutes or more provided the gas is discharged at the

specified rate. Therefore, if your calibration time is estimated at four minutes, you can operate

the zero-point calibration twice.

<Treatment after completion of calibration>
(1) Fully close the stop valve mounted on the calibration gas inlet of the detector.
(2) Remove the tube connecting the detector to the standard gas unit.

WARNING

Store the standard gas unit with the gas cylinder mounted where the ambient temperature does

not exceed 40 °C. Otherwise, the gas cylinder may explode. Store the spare gas cylinders under

the same condition.

8th Edition : Jan.13,2012-00