2.5

ZA8F Flow Setting Unit and

ZR40H Automatic Calibration Unit

2.5.1

ZA8F Flow Setting Unit

This flow setting unit is applied to the reference gas and the calibration gas in a system

configuration (System 2). Used when instrument air is provided.
This unit consists of a flowmeter and flow control valves to control the flow rates of calibration gas

and reference gas.

Standard Specifications
FIowmeter Scale:

Calibration gas; 0.1 to 1.0 l/min. Reference gas; 0.1 to 1.0 l/min.

Construction:

Dust-proof and rainproof construction

Case Material:

SPCC (Cold rolled steel sheet)

Painting:

Baked epoxy resin, Dark-green (Munsell 2.0 GY 3.1/0.5 or equivalent)

Tube Connections:

Rc1/4 or 1/4FNPT

Reference Gas pressure: Clean air supply of sample gas pressure plus approx. 50 kPa G

(or sample gas pressure plus approx. 150 kPa G when a check valve

is used). Pressure at inlet of the Flow Setting Unit.(Maximum 300 kPaG)

Air Consumption:

Approx. 1.5 l/min

Weight:

Approx. 2.3 kg

Calibration gas (zero gas, span gas) Consumption: Approx. 0.7 l/min (at calibration time only)

NOTE

Use instrument air for span calibration gas, if no instrument air is available, contact YOKOGAWA.

Model and Codes

Model

Suffix code Option code

Description

ZA8F

- - - - - - - -

- - - - - - - - -

Flow setting unit

Joint

-J

-A

- - - - - - - - -

- - - - - - - - -

Rc 1/4

With 1/4 NPT adapter