<10. Other Functions>

10-6

IM 11M12A01-02E

0.0 vol%O

2

25.0 vol%O

2

10 min./div

12.3 vol%O

2

60 data

Preset upper limit

Preset lower limit

Time per division,

calculated by the set

sampling interval

Currently measured

value

Sampling period

F10.5E.ai

Figure 10.5 Plotting Graph for Sampling Interval

10.1.2.3 Setting Upper and Lower Limit Values on Graph

Set upper- and lower-limit values on the graph in the following procedure:
Press “Upper limit” in the Trend graph display. The numeric-data entry key appears.

Enter the upper-limit value. Also enter the lower-limit value in the same way. The allowable

settings for both upper-limit and lower-limit values range from 0 to 100 vol%O

2

.

10.1.2.4 Default Setting

When the analyzer is delivered, or if data are initialized, the set data are by default, as shown in

Table 10.3.

Table 10.3 Default Values for Graph Setting

Item Default

Value

Parameter

Oxygen concentration

Sampling interval

30 seconds

Upper limit

25 vol%O

2

Lower limit

0 vol%O

2

NOTE

If a rapid change in the measured value occurs during sampling, no sampled data are plotted on

the graph. Use the graph indication tentatively. Check the output current for accurate data.