2 system 2, System 2 -2 – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 16

Advertising
background image

<1. Overview>

1-2

IM 11M12A01-02E

8th Edition : Jan.13,2012-00

NOTE

• As this system uses ambient air for the reference gas, measuring accuracy will be affected

by the installation location.

• A needle (stop) valve should be connected to the calibration gas inlet of the detector.

The valve should be fully closed unless calibration is in progress.

1.1.2

System 2

This system is for monitoring and controlling oxygen concentration in the combustion gases of a

large-size boiler or heating furnace. Clean (dry) air (21 vol% O

2

) is used as the reference gas and

the span gas for calibration. Zero gas is fed in from a cylinder during calibration. The gas flow is

controlled by the ZA8F flow setting unit (for manual valve operation).

F1-2E.ai

ZR402G Converter

Check valve

or Stop Valve

ZA8F Flow Setting Unit

EXA ZR402G

ZR22G Separate type

Zirconia Oxygen Analyzer, Detector

100 to

240 V AC

Air Set

Instrument air

Calibration gas

Reference

gas

Flowmeter

Needle

valve

Calibration gas

unit case

Zero gas

cylinder

Span gas

(Same as Zero gas calibration unit)

Contact input

Analog output, contact output

(Digital output HART)

Signal

(6-core shield cable)

Heater(2-core)

~

Pressure

reducing

valve

Figure1.2