6.1.2 Reagent Management

Use the Reagent management screen (Reagents menu, Management) to set the default

number of cassettes per run and default carryover setting per run, to turn wax cleaning on and off,

and to set wax standby and melting temperatures. Supervisor access is required.
A range of other configuration settings can be viewed but not changed – contact customer support

to change any of these. The screen also displays the state of each retort and the last reagent in

each retort.
Controls on the Reagent management screen are described below, grouped by section.

General

Conc. Management — (view only) the method the system uses to assign concentration

values to reagent bottles (see 5.1.2 Concentration Management).

Prompt for number of cassettes — (view only) under the default “enabled” setting Peloris

prompts you to enter the number of cassettes before each run. When the function is disabled,

the system assumes that each run processes the default number of cassettes (see below).

Default number of cassettes — set the default number of cassettes for each processing

run. If Prompt for number of cassettes is enabled the prompt dialog box opens before

each run showing this number of cassettes to accept or change. If Prompt for number of

cassettes is disabled then the system uses the number of cassettes entered here for all

processing runs. In this case be sure to enter an accurate average value – the number is used

by the reagent management system to calculate reagent concentrations.

Default carryover — the default carryover setting used for new protocols until a new value is

set (see 4.1.5 Carryover Setting).

Show conc. — (view only) hides or shows the current wax and bottle concentrations on the

Status screen.

Reagent Threshold Check

The four buttons in the Reagent threshold check section of the Reagent management screen

show which types of reagent thresholds the system is using (see 5.1.3 Thresholds). By default all

four threshold types are enabled. Contact your customer support representative to change any of

these settings.

By cassettes — the system tracks the number of cassettes processed by the reagent in each

station. If the reagent type has a cassette threshold set you are alerted when the station goes

beyond that threshold.

By cycles — the system tracks the number of processing runs performed by the reagent in

each station. If the reagent type has a cycle threshold set you are alerted when the station

goes beyond that threshold.

By days — the system tracks the number of days the reagent in each station has been loaded.

If the reagent type has a days threshold set you are alerted when the station goes beyond that

threshold.

By conc. — the system tracks the concentration of reagent in each station. If the reagent type

has a concentration threshold set you are alerted when the station goes beyond that threshold.