8.2.4 Xylene-free Protocols

All the xylene-free protocols below are designed for use with the xylene-free bottle configuration in

8.3 Station Configurations.

Xylene-free 1 Hour

Pre-defined protocol “Factory 1 hr Xylene Free“

Carryover setting in pre-defined protocol: 100

Xylene-free 2 Hour

Pre-defined protocol “Factory 2 hr Xylene Free“

Carryover setting in pre-defined protocol: 75

Step Reagent type

Time (min) Temp (°C) P/V

Stirrer

Drip time

(s)

1

Formalin

1

Ambient

Ambient

Medium

10

2

85% Ethanol

1

Ambient

Ambient

Medium

10

3

85% Ethanol

6

55

Ambient

Medium

10

4

80/20 Ethanol/IPA

1

Ambient

Ambient

Medium

10

5

80/20 Ethanol/IPA

6

55

Ambient

Medium

10

6

IPA

1

Ambient

Ambient

Medium

10

7

IPA

1

Ambient

Ambient

Medium

10

8

IPA

12

55

Ambient

Medium

10

9

Wax

20

85

Vacuum

Medium

10

10

Wax

5

85

Vacuum

Medium

10

11

Wax

1

65

Vacuum

Medium

10

Processing time

1:19:00

Step Reagent type

Time (min) Temp (°C) P/V

Stirrer

Drip time

(s)

1

Formalin

1

Ambient

Ambient

Medium

10

2

85% Ethanol

1

Ambient

Ambient

Medium

10

3

85% Ethanol

12

55

Ambient

Medium

10

4

80/20 Ethanol/IPA

1

Ambient

Ambient

Medium

10

5

80/20 Ethanol/IPA

25

55

Ambient

Medium

10

6

IPA

1

Ambient

Ambient

Medium

10

7

IPA

1

Ambient

Ambient

Medium

10

8

IPA

25

55

Ambient

Medium

10

9

Wax

25

85

Vacuum

Medium

10

10

Wax

10

85

Vacuum

Medium

10

11

Wax

5

65

Vacuum

Medium

10

Processing time

2:11:00