1 manual operations, Manual operations, 1 manual opera – Leica Biosystems PELORIS_PELORIS II User Manual

Page 99: Tions, See 6.1.1 manual, Operations), Retort temperature range

Advertising
background image

Settings & Ancillary Operations

Leica PELORIS™ User Manual Rev K © Leica Biosystems Melbourne Pty Ltd 2011

99

6.1.1 Manual Operations

The Manual operations screen (Reagents menu, Manual operations) allows you to manually

control the instrument. The screen displays the current instrument status including station states,

the retort states and the retort condition. You control the instrument by selecting instrument

elements from the status area and control functions from the panels on either side.

All users can perform manual operations functions.

You can fill a retort from any station and then set conditions in the retort: temperature, pressure

and stirrer. If intending to move wax into a retort preheat the wax path by turning on the wax

heater. Use the wax vent function to neutralize the pressure in the wax bath so the wax bath lids

can be opened easily and safely.

You cannot override a running protocol from the Manual operations screen and you cannot fill or

drain a retort that has a loaded protocol.

Figure 58 shows the Manual operations screen with function buttons explained:

Figure 58. Manual operations screen

Retort Temperature Range

The retort temperature range is limited to the following values.

Reagent — 35 ºC minimum to 85 ºC maximum (restricted to 65 ºC maximum when running

tissue processing protocols).

Wax — (Wax melting point + 2 ºC) minimum to 85 ºC maximum (restricted to 77 ºC when

running standard protocols).
View the current wax melting point temperature on the Reagent management screen.

Fill/Drain retort:

fill or drain the retort

from or to the selected

station

Stirrer: set the retort

stirrer to high, medium

or low speed, or off

Temp.:

set the retort

temperature

Retort A controls

Wax heater:

turn on/off heaters for

wax path

Pressure:

set the pressure in the

retort – ambient,

vacuum, pressure, or

pressure/vacuum cycle

Abandon:

stop a fill or drain

operation

Retort B controls

Vent wax:

release pressure or

vacuum in wax chambers

Advertising