Troubleshooting

Leica PELORIS™ User Manual Rev K © Leica Biosystems Melbourne Pty Ltd 2011

143

9.2.3 Under- or Over-processed Tissue – Reagents

Are the

reagents

physically

correct?

Check:

·

reagent concentrations with a

hydrometer

·

the reagent change and final reagent

thresholds

·

bottles are in their correct stations

·

the reagents in the bottles are correct

·

reagent storage conditions

·

reagent threshold check method

Are they all correct?

Check:

·

reagents are clean

Factors that could affect reagent purity are:

·

recycled reagents

·

running reprocessing protocols

·

failing to clean the retorts before or after
remote drain or fill operations

Stop

Well processed tissue

Discard incorrect or dirty

reagents. Rinse reagent

bottles and fill and drain

retort with fresh reagent to

clean reagent lines.

No

No

Are the

reagents clean?

Under/over-processed

tissue -reagents

Yes

No

Yes

Yes

No

Yes

Is the wax

contaminated

or does it smell like

formalin?

No

Open a service call to have Leica

Microsystems test the instrument and
repair if required.

Is the instrument functioning correctly?

Retest. Is the

tissue processed

correctly?

Yes