Xylene 12 Hour

Pre-defined protocol “Factory 12hr Xylene Standard“

Carryover setting in pre-defined protocol: 25

Step Reagent type

Reagent

group

Time

(min)

Temp

(°C)

P/V

Stirrer

Drip time

(s)

1

Formalin

Fixatives

44

45

Ambient

Medium

10

2

Ethanol

Dehydrants

30

45

Ambient

Medium

10

3

Ethanol

Dehydrants

30

45

Ambient

Medium

10

4

Ethanol

Dehydrants

30

45

Ambient

Medium

10

5

Ethanol

Dehydrants

30

45

Ambient

Medium

10

6

Ethanol

Dehydrants

60

45

Ambient

Medium

10

7

Ethanol

Dehydrants

90

45

Ambient

Medium

10

8

Xylene

Clearers

45

45

Ambient

Medium

10

9

Xylene

Clearers

45

45

Ambient

Medium

10

10

Xylene

Clearers

90

45

Ambient

Medium

10

11

Paraffin wax

Wax

60

65

Vacuum

Medium

10

12

Paraffin wax

Wax

60

65

Vacuum

Medium

10

13

Paraffin wax

Wax

80

65

Vacuum

Medium

10

Processing time

12:02:00