5.4.3 Replacing Reagent – Remote Fill and Drain

You can drain and then refill reagent bottles without removing the bottles from the instrument. The

process drains each bottle to a retort then drains the retort to the remote fill/drain line. The reverse

procedure is used for the fill.

If you routinely use remote drain and fill do not forget to check if bottles need cleaning, once a

week.

Before you begin a remote fill or drain ensure there is a retort available:

the retort must not have a protocol loaded or running;

The retort must be clean or empty;

The retort residue (if any) must be compatible with the reagent in the bottle(s).

Follow the instructions below to replace reagent in a bottle.

Drain

1. Connect the remote fill/drain hose and place the end in a suitable container (see 5.4.2 Remote

Fill/Drain Connections).

2. From the Remote fill/drain screen (Reagents menu, Remote fill/drain), select:

The retort to use

The Remote icon

The bottle(s) to drain (multiple bottles must all have reagent in the same group).

Figure 54. Remote fill/drain setup for draining bottles

3. Tap Drain from bottle(s) to begin the drain.

Select the retort to use

Select the bottle or bottles to

drain

Select Remote icon

When bottle, retort and

Remote are selected tap

Drain from bottle(s) to

start the drain