background image

Reagent Setup

Leica PELORIS™ User Manual Rev K © Leica Biosystems Melbourne Pty Ltd 2011

94

5.4.4 Replacing Reagent – Manual

To replace bottles manually remove the bottles and dispose of the old reagent following your

laboratory’s standard procedures. Clean the bottle if necessary, and then fill with fresh reagent.

Load back on to the instrument.

When the bottle is reinserted the Inserted bottle configuration dialog box opens, showing the

reagent type, concentration and use history for either:

the bottle when it was removed, or

the bottle as configured in the Reagent stations screen, if you assigned a new reagent type

to the station after the bottle was removed (see Assigning New Reagents to Stations).

Figure 56. Inserted bottle configuration dialog box

The bottle state is set to Unknown in the dialog box.

Tap the table row and select one of the options at the bottom of the dialog box:

Changed — select if you replaced the old reagent with fresh reagent.
With this option the software prompts you to accept the default concentration for the reagent

in the station. If you know the concentration to be other than the default you can enter a

different concentration.
The use history values are all set to zero.
If you changed the reagent type in the bottle you cannot update it in the software from this

dialog box. Tap No change and then remove the bottle again. Change the reagent assigned to

the bottle in the Reagent stations screen (see Assigning New Reagents to Stations), then

reinsert the bottle.

Topped off/up — select if you did not change all the reagent but added a small amount of

fresh reagent of the same type to bring up its level in the bottle.
With this option the bottle state changes to Full. The concentration and use history details do

not change.

No change — select if you made no changes to the reagent in the bottle at all.

Tap in the table row to

select the bottle, then

select an option button

below

No change:

select if you did not do

anything to the

reagent in the bottle

Topped off/up:

select if you added a

small amount of reagent

to the bottle to bring up

its level

Changed:

select ONLY if you replaced

the bottle contents with

fresh reagent