N use” (see setting station states, Changing reagent concentration, Setting station states – Leica Biosystems PELORIS_PELORIS II User Manual

Page 87

Advertising
background image

Reagent Setup

Leica PELORIS™ User Manual Rev K © Leica Biosystems Melbourne Pty Ltd 2011

87

Changing Reagent Concentration

You can set a station’s “calculated” concentration value. Tap the station’s Conc. cell. Enter the new

concentration in the keypad.

Setting Station States

All users can change station states (Full, Empty, In use, Dry, and Not molten) in the Reagent

stations screen.

For reagent bottles you should not normally need to do this – you update the bottle states when

you remove and replace bottles and then the software tracks the state automatically. Change bottle

station states only if the wrong state is displayed or you want to make a full bottle unavailable for

use (by setting it to In use).

For wax chambers you must change the station state in the Reagent stations screen as a normal

part of wax replacement.

To change a station state tap the station’s State cell and select the appropriate icon in the dialog

box.

See 3.3.1 Status Area for detailed information on reagent station states.

WARNING
Do not alter the concentration of a used reagent unless you are able to verify the actual
concentration. If the concentration is incorrect a reduction in tissue processing quality or
damage to the tissue sample may result.

Figure 49. Station state dialog box for reagent stations (L) and wax stations (R)

WARNING
Ensure you set the station state to the actual condition of the station. An incorrect reagent
station state may cause fluid leaks or abandoned processing runs.

Advertising