9.2.2 Over-processed Tissue – Instrument Setup

Over-processed tissue

(cooked/ crispy/ burnt)

Yes

Was tissue

allowed to dry

out?

Check:

·

How was the tissue submitted to the lab-
in formalin?

·

Was tissue in solution between grossing
and processing?

·

Was the protocol in dehydrant for too
long?

·

Was the protocol in solvent for too long?

·

Was tissue left in the retort in wax for too
long?

·

Was the tissue left in the embedding
instrument too long?

Yes

No

No

No

No

No

Yes

No

Yes

No

Stop

Well Processed Tissue

Yes

Yes

Yes

Go to 9.2.3 Under- or

Over-processed Tissue – Reagents

No

Are any error

codes present?

If a user warning, fix the

issue. If a service warning,
contact service

Retest. Is the

tissue processed

correctly?

Was the correct

protocol used?

Check:

·

tissue size

·

step times

·

pressure and vacuum

·

temperatures

·

drip times

Are the settings appropriate?
Optimise the reagent using the pre-defined

protocols as a guideline.

Was the

correct carryover

value, number of

cassettes,

entered?

Enter correct values for Carryover and

cassettes for the reagent management
system to work correctly.

Retest. Is the

tissue processed

correctly?

Retest. Is the

tissue processed

correctly?

Retest. Is the

tissue processed

correctly?

Is the lab

workflow

appropriate?

Check:

·

Transfer fluid (formal saline?)

·

Fixation time

·

Water introduced after processing?

·

Quality of embedding wax

Are these all correct?

No

Retest. Is the

tissue processed

correctly?

Yes

Yes

Yes

No