Settings & Ancillary Operations

Leica PELORIS™ User Manual Rev K © Leica Biosystems Melbourne Pty Ltd 2011

107

Toggle the Show by time/Show by frequency button to switch between the two views.

Tap the Filter button to open the Events log filter dialog box where you set filters to select

the types of events to view.
Select buttons in the Retorts filters, Severity filters and Alarm filters sections to see the

associated types of events. You must select at least one button in each section to see any

events. E.g. if you select Retort A in Retorts filters, Information in Severity filters and

Non-alarmed events in Alarm filters you will see all information events for retort A that do

not raise an alarm.
The Run log entry button in the filter dialog box is only enabled for service personnel.

To create a user event select Create user event to open the on-screen keyboard. Type in the

information you want to record and tap Enter to add the message as an event. The event

severity will be “Information”.

To create an incident report file for a specific event (to send to your service representative)

select the event row in the table and tap Incident report (see Transferring Incident Report

Files in 6.2.7 File Transfer).

Event severity

There are four event severity levels and each level has a color code as shown in the following table.

6.2.5 Alarms

Peloris uses alarms to alert users when events of certain types occur. These events are mostly for

error conditions that require quick intervention, but alarms also activate for some normal conditions

that it is useful to be informed about, such as when a protocol has finished.

Severity level

Description

Color code

Information

A normal event that requires a response (e.g. a protocol

successfully completed) or an unusual event that has no

detrimental effect (e.g. a user abandoned protocol).

Dark green

Warning

An error or potential error that does not stop processing, or a

request for user action. For example, an out-of-threshold

reagent used in a protocol.

Orange

Error

An error that causes an operation to abandon (e.g. a protocol

abandons because there is no station available).

Red

Critical error

An error that makes part of the instrument (e.g. one retort) or

the entire instrument unusable.

Blue