9.2.4 Poor Processing – Incorrect Protocol

Poor processing – incorrect

protocol

Was

the original

protocol too

long?

Tissue is over-
processed.
Reprocessing will not
help.

Was

the original

protocol too

short?

Tissue is under processed. Reprocess. NB: Do not use the Peloris clean
cycle to reprocess the tissue as the drying step will cause the tissue to dry
out.

Is it

large robust

tissue?

Is it fatty

tissue?

Try a longer protocol with
less heat and agitation.

Try a longer, protocol with
more heat and efficient
agitation.

Try a defatting step of 50/50
xylene and ethanol. Put step
between ethanols and xylene,
ensuring that there is a 100%
ethanol step after defat,
before xylene.

Is tissue

processed

adequately?

Yes

Yes

Yes

Yes

No

Was wax

infiltration

incomplete?

If tissue is dehydrated and cleared
there is no need for reprocessing.
Re-infiltrate tissue with wax only.

Yes

No

Adequately processed

tissue

Yes

Try:

·

soaking the block in a softening agent, such
as fabric softener, before cutting.

·

Cutting thinner sections.

·

Warming the block slightly before sectioning

No

Contact customer

support

Is it xylene

processing?

No

No

Is it

large delicate

tissue?

No

No

Yes

No